SECTION IV. CALIBRATION PROCESS (SECONDARY TRANSFER ONLY)
Calibration Factors

Calibration Procedure For Power Sensors And Thermistor Mounts 10 Mhz To 40 Ghz (General) Page Navigation
  12    13    14    15    16  17  18    19    20    21    22  
img
. .
TB 9-6625-1932-24
i. Set power meter No. 1 CAL FACTOR % control to value listed on TI body for lowest
test frequency.
j.  Set power meter No. 2 CAL FACTOR % control to value listed on standard
thermistor mount body for lowest test frequency.
k. Connect equipment as shown in figure 8, CONNECTION A, and allow 1 hour for
warm-up.
Figure 8. Calibration factor equipment setup.
17