Calibration Factors
APPENDIX A. TEST INSTRUMENT SPECIFICATIONS

Calibration Procedure For Power Sensors And Thermistor Mounts 10 Mhz To 40 Ghz (General) Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
. . .
TB 9-6625-1932-24
NOTE
Do not disturb synthesized signal generator output controls
while performing (18) below.
(18) Record power meter No. 1 indication as P2.
(19) Compare power meter No. 2 (P1) value recorded in (14) above with power meter
No. 1 (P2) level recorded in (18) above. Differences in power levels will not exceed the
tolerance listed in Appendix A.
NOTE
Perform b below for Hewlett-Packard, Models 478A and 8478B
that fail (19) above.
(20) Repeat technique of (1) through (19) above for a test frequency near the center
and one at the upper end to TI response range or at specified frequencies requested by
activity submitting TI for calibration.
b. Adjustments. Perform 8 b above.
13. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.
19/(20 Blank)