TB 9-6625-1942-24
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Range: 0 to 1050 V ac
Fluke, Model 5720A (5720A) (p/o
MIS-35947); w amplifier, Fluke
0.0325%
5725A/AR (5725A/AR)
Accuracy:
Range: 0 to 1000 V dc
0.00375%
Accuracy
Range: 0 to 1 M
Accuracy: 0.00425%
10 M
Accuracy: 0.0105% (0.015%)
DC VOLTAGE DIVIDER
Range: 10:1
ESI Model, RV722 (RV722)
Accuracy:
0.00175%
Range: 0 to 10 V dc
Agilent, Model 3458A
Resolution: 1 V
(3458A)
6. Preliminary Instructions
a. The instruction outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.