TB 9-6625-1943-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin
provides instructions for the
calibration of Frequency Meter, American Machine and Foundry Co., Model PFM 604B,
A and M Instrument, Inc., Model 5009 and Sun Electric Corp., Model M7-110. The
manufacturers' manuals were used as the prime data source in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test
instrument) throughout this bulletin.
a. Model Variations . Variations among models are described in text.
b. Time and Technique . The time required for this calibration is approximately 1
hour, using the dc and low frequency technique.
2. Forms,
Records, and
Reports. Forms, records, and
reports required for
calibration personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Input voltage
105 to 135 V ac
Input frequency
60 and 400 Hz
Meter scales
58 to 62 Hz and 380 to 420 Hz
Resolution
0.1 Hz (58 to 62 Hz) and 1 Hz (380 to 420 Hz)
Meter accuracy:
Center scale
0.1%
Entire scale
0.2%
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity.
The items selected must be verified to perform satisfactorily prior to use and must bear
evidence of current calibration. The equipment must meet or exceed the minimum use
5. Accessories Required. The
accessories required for this
calibration are
common
usage accessories issued as indicated in paragraph 4 above and are not listed in this
calibration procedure.