TB 9-6625-1943-35
Table 2. Minimum Specifications of
Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
CALIBRATOR
Range: 105 to 135 V ac
John Fluke, Model 5700A/CT (p/o MIS-
35947)
Accuracy: 0.025%
6. Preliminary Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration process. Personnel should become familiar with the entire bulletin before
beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments and additional maintenance information are contained in the
manufacturers' manuals for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance
of this calibration. DEATH ON CONTACT may result if
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Adjust TI mechanical zero (front panel) until each meter indicates first major
division.
b. If necessary, set TI controls as required.