TB 9-6625-1957-24
(4) Repeat technique of (2) (b) and (3) above, using outputs and settings listed in
Test instrument
Calibrator
Indication limits
Range
setting
Min
Max
output
190 A
A
200
189.59 A
190.41 A
1.9 mA
2 mA
1.8959 mA
1.9041 mA
19 mA
20 mA
18.959 mA
19.041 mA
190 mA
200 mA
189.59 mA
190.41 mA
1.9 A
2000 mA
1895.9
mA
1904.1
mA
(5) Reduce calibrator output to minimum.
b. Adjustments. No other adjustments can be made.
a. Performance Check
(1) Connect calibrator OUTPUT HI to TI VOLTS/ input and calibrator OUTPUT
LO to TI LOW input.
(2) Press TI pushbuttons as listed in (a) through (c) below:
(a) k (in).
(b) 20 k range (in).
(c) HI LO (out).
(3) Set calibrator for a 19k
nominal output.
(4) Rotate calibrator knob below EDIT FIELD pushbutton to adjust calibrator
display indication to equal TI indication. If calibrator err display does not indicate within
limits specified in first row of table 11, perform 20 b below.
(5) Repeat technique of (2) (b) and (c), (3) and (4) above, using calibrator outputs
and TI settings and indications listed in table 11. Calibrator err display will indicate
Test instrument
Calibrator
err
HI-LO
display
Nominal
Range
Pushbuttons
indications
Output
pushbuttons
LO
0.166 %
201
k
19
k
HI
0.166 %
20
k
19
k
LO
0.166 %
200
190
LO
0.166 %
2
k
1.9
k
HI
0.166 %
2
k
1.9
k
HI
0.166 %
2002
k
190
k
LO
0.166 %
200
k
190
k
LO
0.316 %
2000
k
1.9
M
HI
0.166 %
2000
k
1.9
M
HI
0.516 %
20
M
19
M
Set calibrator 2 WIRE COMP to on.
1