TB 9-6625-1957-35
b. Adjustments.
(1) Release HI LO pushbutton (HI) and set calibrator for a nominal 19 kΩ output.
(2) Adjust HI Ω R1426 (fig. 2) for a TI indication equal to calibrator's displayed
value for 19 kΩ (R).
(3) Press HI LO pushbutton (LO); then adjust LO Ω R1412 (fig. 2) for a TI
indication equal to calibrator's displayed value for 19 kΩ (R).
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
BALLANTINE, MODEL 10800C, FLUKE, MODEL 80K-40,
AND TEKTRONIX, TYPE 010-0277-00
22. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 22 and
23 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this bulletin. Additional
maintenance information is contained in the manufacturers' manuals for this TI
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a.
Connect digital
multimeter to power module and
connect power module to a
115 V ac source.
b. Pull power module PWR switch to ON and allow at least 15 minutes for
stabilization.