TB 9-6625-1957-35
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Dc voltage:
Fluke, Model 5720A (5700A/EP)
Range: .19 to 1000 V
(p/o MIS-35947), w/ amplifier,
Accuracy: 0.016%
Fluke, 5725A/AR (5725A/AR)
Ac voltage:
Range: 100 mV to 500 V
Accuracy: (%)
40 Hz to 10 kHz
0.162
20 Hz and 20 kHz
0.26
Range: 190 Ω to 19 MΩ
Accuracy: (%)
190 Ω through 1.9 MΩ
0.042
19 MΩ
0.128 dBm (6000
reference) (Types DM501A and DM501AOPT02)
Range: -20 to +45 dBm
Accuracy: 0.125 dBm (l.4%)
Dc current:
Range: 190 A to 1.9 A
Accuracy:
0.054%
Dc voltage:
Fluke, Model 8840A/AF-05/09
(AN/GSM-64D)
Range: -11.4 to 12.6 V
Accuracy: 0.5%
TEKTRONIX, TYPE DM 501
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals and/or technical
manuals for the TI.
d. When indications specified in
paragraphs 8 through 11 are not within
tolerance,
perform power supply check prior to making adjustments. After adjustments are made,