Calibration Procedure For Frequency Counter Models 5326A, 536B, And 5326C Page Navigation
  7    8    9    10    11  12
img
. . . .
TB 9-6625-1973-24
b. Adjustments
(1) Adjust A15R10 (fig. 1) for a +16.5 V indication on multimeter (R).
(2) Adjust A15Rl3 (fig. 1) for a -16.5 V indication on multimeter (R).
15. Final Procedure
a. Deenergize and disconnect all equipment and replace TI within protective cover.
b. Annotate and affix DA label/form in accordance with TB 750-25.
13/(14 Blank)