TB 9-6625-1976-35
NOTE
NULL function preprogrammed, you will have to access the
MATH function through the MENU command and enable
MATH NULL.
(6) Disconnect calibrator and connect TI OUTPUT to multimeter Input (2 WIRE)
terminals.
(7) Set TI as listed in (a) through (g) below:
(c) VOLTAGE RANGE switch to 1 V.
(d) VOLTAGE dials to 1.000000 V.
(e) VOLTAGE ERROR-% switch to OFF.
(f) SENSE switch to INT.
(g) MODE switch to OPER.
(8) Allow multimeter to take at last three readings (samples). Multimeter will
indicate 0 V (0.00122 V); if not, perform steps 1 and 2 of table 6 and b (1) through (5)
below.
(9) Repeat technique of (2) through (8) above for remaining steps listed in table 6,
using settings and values specified.
Table 6. Voltage Accuracy (Transfer Level)
Multimeter
Calibrator output
Test instrument
indications
(V)
FREQUENCY
FREQUENCY
VOLTAGE
Voltage
0 (null)
RANGE
RANGE-Hz
dials
dials
volts
Volts
switch (V)
switch
(Hz)
(V)
Step
(V)
Frequency
Adjustment
(V)
1
1.0
1.0 kHz
1K
1.0000 K
10
1.00000
0.00140
b(1)
2
10
1.0 kHz
1.0000 K
10
10.00000
0.01220
b(2) and (3)
1 K
3
100
1.0 kHz
1.0000 K
100
100.0000
0.1220
b(4)
1 K
4
10
1.0 MHz
1 M
1.0000 M
10
10.00000
0.12301
b(5) thru
(7)
5
1.0
1.0 MHz
1 M
1.0000 M
10
1.00000
0.01502
6
1.0
1.0 MHz
1 M
1.0000 M
1
1.00000
0.01230
b(8)
12