TB 9-6625-1976-35
Table 6. Voltage Accuracy (Transfer Level) - Continued
Multimeter
Calibrator output
Test instrument
indications
(V)
FREQUENCY
FREQUENCY
VOLTAGE
Voltage
0 (null)
RANGE
RANGE-Hz
dials
dials
Volts
volts
switch (V)
switch
(Hz)
(V)
Step
(V)
Adjustment
(V)
7
10
1.0 MHz
1 M
1.0000 M
100
10.0000
0.150
b(9)
8
10
0.30 MHz
1 M
.300 M
100
10.0000
0.150
9
10
20 Hz
100
20
100
10.0000
0.0170
10
10
100 kHz
100
100
10
10.00000
0.01220
b(1)
11
10
10 kHz
10 K
10 K
10
10.00000
0.01220
12
1.0
100 kHz
100 K
100.00 K
1
1.00000
0.01205
13
1.0
50 kHz
100 K
50.00 K
1
1.00000
0.00125
14
1.0
10 kHz
10 K
10 K
1
1.00000
0.01220
15
1.0
20 Hz
100
20
1
1.00000
0.01220
16
1.0
1.0 kHz
1K
1.0000 K
1
1.00000
0.001223
(Note change in
calibrator output)
FREQUENCY
FREQUENCY
VOLTAGE
Voltage
0 (null)
RANGE
RANGE-Hz
dials
dials
Volts
volts
switch (mV)
switch
(Hz)
(mV)
Step
(mV)
(mV)
17
100
1.0 kHz
1K
1.0000 K
100
100.0000
0.13
18
10
1.0 kHz
1K
1.0000 K
10
10.0000
0.022
NOTE: The following % of error data will not be utilized if indications in steps 4 and 5 are within limits:
1Record multimeter indication, then convert to % of error by multiplying value by 10. If out of tolerance, perform step 5
above prior to performing b (7) through (9) below.
2Record multimeter indication, then convert to % of error by multiplying value by 100.
3If indication is within limits, adjust TI VOLTAGE dials for the best null indication on multimeter. Record resulting TI
VOLTAGE dial setting for use in (15) below.
13