Table B-99. AC, AC+DC Current.
Table B-104. Calibration Description.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  48    49    50    51    52  53  54    55    56    57    58  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Amprobe Instruments, Model 38XR-A Continued
Table B-101. Capacitance.
Capacitance standard
Test instrument
output
Applied
Range
Limits (F)
(F)
(F)
Min
Max
36 n
40 n
34.82 n
37.18 n
360 n
400 n
348.7  n
371.3  n
3μ
4μ
2.905 μ
3.095 μ
36 μ
40 μ
34.87 μ
37.13 μ
360 μ
400 μ
348.2  μ
371.8  μ
Table B-102. Frequency.
Function generator output
Test instrument
Applied
Limits (Hz)
(V ac)
Min
Max
1.6 @ 90 Hz
89.86
90.14
1.6 @ 900 Hz
898.6
901.4
1.6 @ 9 kHz
8.986 k
9.014 k
1.6 @ 90 kHz
89.86 k
90.14 k
2.1 @ 1 MHz
0.994 M
1.006 M
2.1 @ 10 MHz
9.94 M
10.06 M
Table B-103. Temperature.
Cool/Heat source
Test instrument
Applied
Limits ()
()
Range
Min
Max
0C
C
-4 C
4C
180 C
C
175 C
185 C
500 C
C
488 C
512 C
1
32 F
F
23 F
41 F
360 F
F
350 F
370 F
900 F
F
878 F
922 F
Set TI Function/Range switch to F.
1