TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Meterman, Model 5XP - Continued
Table B-107. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
180 μ
200 μ
177.2 μ
182.8 μ
1.8 m
2m
1.772 m
1.828 m
18 m
20 m
17.72 m
18.28 m
180 m
200 m
177.2 m
182.8 m
Table B-108. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
180 μ @ 45 Hz
200 μ
175.9 μ
184.1 μ
180 μ @ 500 Hz
200 μ
175.9 μ
184.1 μ
1.8 m @ 45 Hz
2m
1.759 m
1.841 m
1.8 m @ 500 Hz
2m
1.759 m
1.841 m
18 m @ 45 Hz
20 m
17.59 m
18.41 m
18 m @ 500 Hz
20 m
17.59 m
18.41 m
180 m @ 45 Hz
200 m
175.9 m
184.1 m
180 m @ 500 Hz
200 m
175.9 m
184.1 m
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
200
-1.211% (187.7)
+1.211% (192.3)
1.9 k
2k
-1.211% (1.877 k)
+1.211% (1.923 k)
19 k
20 k
-1.211% (18.77 k)
+1.211% (19.23 k)
190 k
200 k
-1.211% (187.7 k)
+1.211% (192.3 k)
1.9 M
2M
-1.737% (1.867 M)
+1.737% (1.933 M)
19 M
20 M
-3.263% (18.38 M)
+3.263% (19.62 M)
B-45