TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Meterman, Model 5XP
Test instrument parameters
Performance specifications
DC VOLTAGE
Range: 0 to 1000 V dc
Accuracy: (1.0% rdg + 1 dgt)
AC VOLTAGE
Range: 0 to 750 V ac, 45 to 500 Hz
Accuracy: (1.5% rdg + 5 dgts)
DC CURRENT
Range: 0 to 200 mA dc
Accuracy: (1.5% rdg + 1 dgt)
AC CURRENT
Range: 0 to 200 mA ac, 45 to 500 Hz
Accuracy: (2.0% rdg + 5 dgts)
Range: 0 to 20 MΩ
Accuracy: 200 Ω to 200 kΩ rngs, (1% rdg + 4 dgts);
2 MΩ rng, (1.5% rdg + 4 dgts);
20 MΩ rng, (3% rdg + 5 dgts)
Table B-105. DC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V dc)
(V dc)
(V dc)
Min
Max
180 m
200 m
178.1 m
181.9 m
1.8
2
1.781
1.819
6
20
5.93
6.07
14
20
13.85
14.15
18
20
17.81
18.19
-18
20
-18.19
-17.81
180
200
178.1
181.9
900
1000
890
910
Table B-106. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
180 m @ 45 Hz
200 m
176.8 m
183.2 m
180 m @ 500 Hz
200 m
176.8 m
183.2 m
1.8 @ 45 Hz
2
1.768
1.832
1.8 @ 500 Hz
2
1.768
1.832
18 @ 45 Hz
20
17.68
18.32
18 @ 500 Hz
20
17.68
18.32
180 @ 45 Hz
200
176.8
183.2
180 @ 500 Hz
200
176.8
183.2
675 @ 45 Hz
750
660
690
675 @ 500 Hz
750
660
690
B-44