TB 9-6625-1996-35
(4)
SAMPLE RATE control fully ccw.
(5)
CHANNEL A and B LEVEL controls to PRESET.
(6)
CHANNEL A and B SLOPE switches to + (positive).
CHANNEL A and
B input
impedance switches to
50 Ω.
(7)
(8)
CHANNEL A and B ATTEN switches to X1.
(9)
CHANNEL A and B AC-DC switches to DC.
(10)
CHECK-COM A-SEP switch to SEP.
a. Performance Check
(1) Connect a time/frequency workstation 1 MHz output to frequency difference meter
REF INPUT.
(2) Connect TI FREQ STD OUTPUT 10 MHz (rear panel) to frequency difference
meter SIG INPUT.
(3) Adjust A18 FREQ ADJ (fig. 1) for minimum difference indication on frequency
difference meter.
(4) Allow at least 24 hours for stabilization. Frequency difference meter indication
will remain within 5 parts in 10-10.
(5) Adjust autotransformer output to 105 V and allow 15 minutes for stabilization.
Verify that TI oscillator drift is less than 1 part in 10-10.
(6) Adjust autotransformer output to 125 V and allow 15 minutes for stabilization.
Verify that TI oscillator drift is less than 1 part in 10-10 .
(7) Adjust autotransformer output to 115 V.
b. Adjustments. No further adjustments can be made.