TB 9-6625-1996-35
b. Adjustments
NOTE
Remove top cover to make adjustments, then replace top cover
before repeating (a) above.
(1) Adjust A12 pot A (R69) (fig. 3) for a 2.0000 0.0005 V dc multimeter indication
(R).
(2) Adjust A12 pot B (R74) (fig. 3) for a 2.0000 0.0005 V dc multimeter indication
(R).
(3) Adjust A12 pot D (R86) (fig. 3) for a 2.0000 0.0005 V dc multimeter indication
(R).
(4) Adjust A12 pot C (R82) (fig. 3) for a +2.0000 0.0005 V dc multimeter indication
(R).
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
CHANGE 2