TB 9-6625-1997-24
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
IF output
Frequency range: 4.3, 5.6, 5.625, 5.65, 10.0, and 115 MHz
11.5 MHz
Accuracy: 0.10%
2 Vrms to 1 Vrms
Level:
Accuracy: 2.5%
Distortion
<4%
Modulation deviation
Range:
10 to 40 kHz
Accuracy: 2 kHz on 10 kHz range
3 kHz on 20 kHz range
5 kHz on 40 kHz range
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer calibration
Standards Set AN/GSM-286, AN/GSM-287 and AN/GSM-705. Alternate items may be
used by the calibrating activity. The items selected must be verified to perform
satisfactorily prior to use and must bear evidence of current calibration. The equipment
must meet or exceed the minimum use specifications listed in table 2. The accuracies listed
in table 2 provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The accessories listed in table 3 are issued as indicated in
paragraph 4 above and are used in this calibration procedure. When necessary, these items
may be substituted by equivalent items unless specifically prohibited.
Table 2. Minimum Specifications of Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
Boonton, Model 1121
AUDIO ANALYZER
Range:
150 Hz to 1 kHz
(1121)
Accuracy: <1%
Range:
950 Hz to 80 MHz
Fluke, Model PM6681/656
Accuracy: 0.00094%
(PM6681/656)
MEASURING RECEIVER
Range:
8 to 45 kHz
Measuring receiver system
Accuracy: 1%
N5530S consisting of: Spectrum
Analyzer, Agilent Model
Range:
7.5 mV to 1.25 V
E4440A (E4440A), Power meter,
(at 10 MHz)
Agilent Model E4419B
Accuracy: 6.25%
(E4419B), and Sensor module,
Agilent ModelN5532A opt 504
(504)
Range:
8 to 26 V dc
Fluke, Model 8840A/AF-05
(AN/GSM-64D)
Accuracy:
0.1%