TB 9-6625-2002-24
(b) MULTIPLIER X1K pushbutton pressed.
(c) VAR AMPLITUDE /ATTENUATOR control adjusted for a 2 V rms indication on
the audio analyzer.
(3) Press STEP ATTENUATOR 10 pushbutton.
Audio analyzer will indicate
between 0.618 and 0.645 V rms.
(4) Release STEP ATTEN 10 pushbutton and press STEP ATTEN 20 pushbutton.
Audio analyzer will indicate between 0.196 and 0.204 V rms.
(5) Release STEP ATTEN 20 pushbutton and press STEP ATTEN 40 pushbutton.
Audio analyzer will indicate between 0.0196 and 0.0204 V rms.
b. Adjustments. No adjustments can be made.
12. Square Wave Amplitude, Rise and Falltime
a. Performance Check
NOTE
Do not use 50
will result.
600 OHM output to the oscilloscope Vertical 1 input through a
(1) Connect TI
600
feedthrough termination (voltage divider, model 11047A).
(2) Position controls as listed in (a) through (d) below:
(a)
MULTIPLIER X1K pushbutton pressed.
(b)
VAR AMPLITUDE /ATTENUATOR control fully cw.
(c)
STEP ATTEN pushbuttons released.
(d)
(3) Measure square wave amplitude.
Square wave amplitude as indicated on
oscilloscope will be 2.25 to 2.75 V.
(4) Adjust FREQUENCY Hz dial to 5 and press MULTIPLIER X100K pushbutton.
Rise and falltime of displayed square wave will be 50 ns or less.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.