TB 9-6625-2002-35
(13) Remove extender and reinstall TI in power module.
Pull module POWER
switch to ON.
AGC TEST
9. Output
Voltage and Frequency
Response
a. Performance Check
(1) Connect TI ∼ 600 OHM output to true rms voltmeter through a 600 Ω
feedthrough termination (Voltage Divider, Hewlett-Packard, Model 11047A (7911560)).
(2) Position controls as listed in (a) through (d) below:
(a)
(b)
MULTIPLIER X1K pushbutton pressed.
(c)
VAR ATTENUATOR/AMPLITUDE control fully cw.
(d)
STEP ATTEN pushbuttons released.
(3) If true rms voltmeter does not indicate at least 2.5 V rms, perform b below.
(4) Adjust VAR ATTENUATOR/AMPLITUDE control for a 1 V rms indication on
(5) Using FREQUENCY Hz dial and MULTIPLIER pushbuttons vary frequency
from 50 Hz to 500 kHz. True rms voltmeter will indicate between 966 and 1035 mV rms.