SECTION III. CALIBRATION PROCESS
Meter Range

Calibration Procedure For Output Meters Ts-585A/U, Ts-585B/U, Ts-585C/U, And Ts-585D/U Page Navigation
  1    2  3  4    5  
img
. . . . .
TB 9-6625-2003-24
c. Set impedance switches to X1-25 and DECIBELS MULTIPLY BY switch to + 20-100.
d. Connect equipment as shown in figure 1.
Figure 1. Impedance accuracy - equipment setup.
8. Impedance Accuracy
a. Performance Check
(1) Adjust resistance standard for 0 ohms.
(2) Adjust calibrator frequency for 1 kHz and 1.000 V ac indication on multimeter.
(3) Adjust resistance standard until multimeter indicates 0.500 V. Resistance
standard indication will be between 23.75 and 26.25 ohms (23.25 and 26.75 ohms for TS-
585A/U).
(4) Repeat technique of (1) through (3) above for TI switch settings listed in table 3.
Resistance standard indications will be within limits specified.