TB 9-6625-2003-24
Test instrument
( )1
switch settings
Min
Max
X1 30
28.50 (27.90)
31.50 (32.10)
X1 40
38.00 (37.20)
42.00 (42.80)
X1 50
47.50 (46.50)
52.50 (53.50)
X1 60
57.00 (55.80)
63.00 (64.20)
X1 80
76.00 (74.40)
84.00 (85.60)
X1 100
95.00 (93.00)
105.00 (107.00)
X1 125
118.75 (116.25)
131.25 (133.25)
X1 150
142.50 (139.50)
157.50 (160.50)
X1 200
190.00 (186.00)
210.00 (214.00)
X.1 100
9.500 (9.300)
10.50 (10.70)
X10 100
950.0 (930.0)
1050.0 (1070.0)
X100 100
9500 (9300)
10,500 (10,700)
1Indications
in parentheses for TS-585A/U.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect calibrator to TI input terminals.
(2) Adjust calibrator frequency for 1 kHz and output for a full-scale (50) indication
on TI. If calibrator does not indicate between 201.65 and 246.47 V (209.62 and 236.38 for
TS-585A/U, perform b below.
(3) Repeat technique of (2) above for TI switch settings listed in table 4. If
calibrator does not indicate within limits specified, and b was not performed in (2) above,
perform b below.
Test instrument
Calibrator indications (V)
DECIBELS
Meter
TS-585B/U, TS-585C/U.
TS-585A/U
switch
indications
and TS-585D/U
settings
Min
Max
Min
Max
100
40
180.36
220.44
188
212
100
30
156.19
190.91
162.62
184.38
100
20
127.53
155.88
132.34
149.46
1001
10
90.180
110.22
96
106
100
5
63.766
77.937
66.74
75.26
10
50
63.766
77.937
66.74
75.26
12
50
20.165
24.647
21.96
23.63
0.12
50
6.3767
7.7938
6.64
7.52
1Use
24 in. leads below 100 V.
2Decrease
output voltage of calibrator before turning DECIBELS switch to 1 and 0.1.
b. Adjustments
(1) Adjust calibrator for 224.06 V.