TB 9-6625-2020-35
EQUIPMENT REQUIREMENTS
4. Equipment Required.
Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity.
The items selected must be verified to perform satisfactorily prior to use and must bear
evidence of current calibration. The equipment must meet or exceed the minimum use
5. Accessories Required. The
accessories required for this
calibration are
issued as
indicated in paragraph 4 above. Common usage accessories are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: Extender, Tektronix Type 067-0645-02 and Power Module, Tektronix TM515.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Range: 180 mV to 1000 V dc
John Fluke, Model 5700A/CT (p/o
Accuracy: 0.042%
MIS-35947); w/power amplifier,
John Fluke, Model 5725A
Range: 190 Ω to 19 MΩ
(5725A)
Accuracy: 190 Ω 0.151%
1.9 kΩ to 1.9 MΩ 0.138%
19 MΩ 0.263%
Range: 180 A to 1.8 A dc
Accuracy: 0.069%
Range: 0.7746 mV to 500 V ac
Accuracy: 0.264%, 20 Hz and 20 kHz
0.153%, 50 Hz
0.312%, 1 kHz
Range: 0 to 15.76 V dc
John Fluke, Model 8840A/AF-
Accuracy: 0.016%
05/09 (AN/GSM-64D)
6. Preliminary Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration process. Personnel should become familiar with the entire bulletin before
beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by