TB 9-6625-2020-35
(3) Short TI HI and LO. Adjust C315 (fig. 1) until TI indicates .000.
(4) Remove short from TI HI and LO.
(5) Connect calibrator OUTPUT HI and LO to TI INPUT HI and LO.
(6) Set calibrator for a 1.8 V output. Adjust R305 (fig. 1) until TI indicates 1.800 V
(R).
(7) Set calibrator for a 180 mV output.
(8) Set TI range/function switch to 200mV DC VOLTS. Adjust R295 (fig. 1) until
TI indicates 180.0 mV (R).
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI INPUT HI and LO.
(2) Set TI range/function switch to 200K OHMS.
(3) Set calibrator for a 190 kΩ output, EX SNS off then, using output adjustment
controls, set calibrator control display Reading equal to TI indication. If calibrator
control display Error indication is not within 0.5526%, perform b below.
(4) Repeat technique of (2) and (3) above using settings and indications in table 4.