Power Supply

Calibration Procedure For Frequency Counter, Model 5340A And Mircrowave Frequency Counter Td-1225A(V)1/U (Model 5342A/H14), Td-1225A(V)2/U (Model 5342A/H16) And, Models 5342A And 5343A Page Navigation
  12    13    14    15    16  17
img
. . .
TB 9-6625-2021-24
b. Adjustments. Adjust A21R17 (fig. 2) for -5.2 V dc indication on multimeter (R).
21. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.