TB 9-6625-2021-35
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity. The
item selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common Name
Minimum use specifications
(part number)
Range: 105 to 125 V ac
General Radio W10MT3AS3
(7910809) or Ridge, Model
9020A (9020A), or Ridge,
Model 9020F (9020F)
FREQUENCY
Tracor, Model 527E (527E)
Accuracy: <1.25 parts in 1010
DIFFERENCE METER
FUNCTION
(SG-1288/G)
Function: Sinewave, 50Ω
Amplitude: 1 to 200 mVp-p
Range: -14.95 to +15.05 V dc
John Fluke, Model 8840A/AF
(AN/GSM-64D)
Accuracy: 0.08%
Wiltron, Model 68347M
Amplitude: -50 to 13 dBm
TIME/FREQUENCY
Frequency: 1 MHz
Autek Systems Corp., Model
620 (MIS-38946)
WORKSTATION
Accuracy: < 1.0 part in 1011