TB 9-6625-2021-35
SECTION III
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d.
When indications specified in
paragraphs 8 through 11 are not within
tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
e. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Remove protective covers from TI as necessary to make adjustments.
b. Set line voltage SELECTOR switch (rear panel) for 115 V ac and verify that correct
fuse (as labeled on rear panel) is installed.
c. Connect TI to
autotransformer. Connect autotransformer to
115 V ac
power source
and adjust to 115 V ac.
d. Position controls as listed in (1) through (4) below:
(1) RESOLUTION Hz pushbutton to 1.
(2) SAMPLE RATE control fully ccw.
(3) INT-EXT switch (rear panel) to INT.
(4) RANGE switch to 10 Hz - 250 MHz.