• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
SECTION I. IDENTIFICATION AND DESCRIPTION
SECTION III. CALIBRATION PROCESS

Calibration Procedure For Rf Power Meter Models 42B, And 42Bd With Power Detectors Models 41-4A, 41-4B And 41-4E Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . . . .
TB 9-6625-2043-24
Table 1. Calibration Description
Test instrument parameters
Performance specifications
115 V ac 10%, 50 to 400 Hz, 15 W
Power requirements
Frequency range
200 kHz to 12.4 GHz with 41-4B power detector
200 kHz to 18 GHz with 41-4E power detector
200 kHz to 7 GHz with 41-4A power detector
Power range
1 nW (-60 dBm) to 10 mW (+10 dBm) in 7 decade
ranges
Accuracy (model 42B):
200 kHz to 4 GHz:
(0.5% of FS +0.2 dB)
Six highest ranges
Approximates  6.27% at 1 mW
(-50 to +10 dBm)
4 to 8.2 GHz:
(0.5% of FS +0.05 dB/mW +0.3 dB)
Approximates  8.58% at 1 mW
8.2 to 12.4 GHz:
(0.5% of FS +0.05 dB/mW +0.4 dB)
Approximates  10.9% at 1 mW
Lowest range
200 kHz to 4 GHz:
(1.0% of FS +0.4 dB)
(-60 to -50 dBm)
Approximates  10.22%
4 to 8.2 GHz:
(1.0% of FS +0.5 dB)
Approximates  12.54%
8.2 to 12.4 GHz:
(1.0% of FS +0.6 dB)
Approximates  14.86%
Accuracy1 (model 42BD):
0.2 dB
1 digit from 200 kHz to 4 GHz
Six highest ranges
0.3 dB
1 digit from 4 to 8.2 GHz
(-50 to +10 dBm)
0.4 dB
1 digit from 8.2 to 12.4 GHz
0.6 dB
1 digit from 12.4 to 18 GHz
Lowest range
0.4 dB
1 digit from 200 kHz to 4 GHz
(-60 to -50 dBm)
0.5 dB
1 digit from 4 to 8.2 GHz
0.6 dB
1 digit from 8.2 to 12.4 GHz
0.8 dB
1 digit from 12.4 to 18 GHz
the 10 mW (+10 dBm) FS range only, add 0.5 dB to the accuracy statement for frequencies
1On
above 4 GHz.
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287 or AN/GSM-705.  Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-
3


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.