TB 9-6625-2043-24
b. Adjustments
(1) Adjust function/arbitrary generator output for an indication of 7.07 mV on true
(2) Adjust R523 (fig. 1 or 2) for a 10.0 V dc indication on multimeter (R).
a. Performance Check
(1) Zero TI meter as described in paragraph 7 h above.
(2) Set FULL SCALE switch to -30 dBm.
(3) Adjust function/arbitrary generator output for a TI meter indication of -30 dBm
(1.000 W). If true rms voltmeter does not indicate between -29.65 and -30.35 dBm,
perform b below.
(4) Repeat technique of (1) through (3) above for values listed in table 3. If true rms
voltmeter does not indicate within limits specified, perform corresponding adjustments.
Test instrument
indications (dBm)
Adjustments (R)
FULL SCALE
Adjust for
switch positions
Meter
indications
(dBm)
indications
Min
Max
(dBm)
-50
-50 dBm
-49.55
-50.45
R525
-50
(10.00 nW)
-40
-40 dBm
-39.6
-40.4
R524
-40
(100.0 nW)
-20
-20 dBm
-19.7
-20.3
R522
-20
(10.00 W)
-10
-10 dBm
-9.75
-10.25
R521
-10
(100.0 W)
-10
-20 dBm
-19.75
-20.25
R543
-20
(10.0 W)
0
0 dBm
-0.2
+0.2
R520
0
(1.000 mW)
0
-10 dBm
-9.8
-10.2
R548
-10
(.100 mW)
+10
+10 dBm
+9.75
+10.25
R519
+10
(10.00 mW)
+10
0 dBm
-0.25
+0.25
R562
0
(1.00 mW)
b. Adjustments
(1) Adjust function/arbitrary generator output for an indication of -30.00 dBm on
(2) Adjust R408 (fig. 1) or R1405 (fig. 2) for a 30 dBm (1.000 W) indication on TI
meter (R).