Reference Oscillator Frequency Accuracy (Models 436A and 436AOPT9/22)
Figure 6. Power reference level - equipment setup.

Calibration Procedure For Power Meters, Ts-3793 (436A And 436Aopt9/22) Models 435A, 435B, 435Bopt001, And 435Be21 Page Navigation
  12    13    14    15    16  17  18    19  
img
. . . .
TB 9-6625-2050-24
Figure 5. REF FREQ and LEVEL adjustments.
15. Power Reference Level
a. Performance Check
(1) With power meter off, remove thermistor mount from interconnect cable. Adjust
multimeter to measure resistance and connect between VRF connector on rear panel of
power meter and pin 1 on thermistor mount end of power meter interconnect cable.
(2) Round off multimeter indication two decimal places and record this value as
internal bridge resistance R of power meter (approximately 200  ).
(3) Reconnect thermistor mount to power meter interconnect cable and connect
equipment as shown in figure 6.
(4) Press TI LINE pushbutton to ON and POWER REF pushbutton to off. Set
power meter switch to on and wait 30 minutes for thermistor mount to stabilize before
proceeding.
(5) Set power meter RANGE switch to COARSE ZERO and adjust front panel
COARSE ZERO control for a zero meter indication.
(6) Zero power meter on the most sensitive range, then set power meter RANGE
switch to 1 mW
17