SECTION I. IDENTIFICATION AND DESCRIPTION
Table 2. Minimum Specifications of Equipment Required

Calibration Procedure For Signal Generators Sg-543U, Sg-543B/U, And Sg-543C/U; And, Models 204B And 204C Page Navigation
  1    2  3  4    5    6    7    8  
img
. . .
TB 9-6625-2057-24
Table 1. Calibration Description
Test instrument parameters
Performance specifications
SG-543B/U and SG-543C/U and Hewlett-Packard, Model 204C
Frequency
Range: 5 Hz to 1.2 MHz
Accuracy:  3%
3%
Stability:
Frequency response
(into 600  )
Normal mode
Accuracy: +5 to -1%, 5 Hz to 100 Hz
0.5%, 100 Hz to 300 kHz
1%, 300 kHz to 1.2 MHz
Low distortion
Accuracy: +5 to -1%, 5 Hz to 100 Hz
0.5%, 100 Hz to 300 kHz
mode
1%, 300 kHz to 1.2 MHz
D
Output voltage
Range: 2.5 V ac into 600
Accuracy:  As indicated in figure 11
istortion
(into 600  )
SG-543/U and Hewlett-Packard, Model 204B
Frequency
Range: 5 Hz to 560 kHz
Accuracy:  3%
3%
Stability:
D
3% into 600
Frequency response
Accuracy:
Less than 1%1
istortion
1
Output voltage
Range: 2.5 V ac into 600
Not calibrated above 100 kHz.
Figure 1. Frequency - Hz.
3