Power Supply

Calibration Procedure For Signal Generators Sg-543U, Sg-543B/U, And Sg-543C/U; And, Models 204B And 204C Page Navigation
  10    11    12    13    14  15
img
. . .
TB 9-6625-2057-24
Figure 5. Test point location.
NOTE
If TI contains battery pack power supply and tolerance is not
met, it may be necessary to recharge or replace battery pack. If
batteries are rechargeable, allow 20 minutes for recharge.
b. Adjustments. No further adjustments can be made.
20. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
15/(16 Blank)