TB 9-6625-2059-35
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Range: 3.2, 8, 150, 600Ω and HI-Z (1 MΩ)
Ac load
Accuracy: 10%
Range: 0.1 to 300 V in 8 ranges
Ac volts
Accuracy: 5% FS
Frequency: 50 Hz to 20 kHz
Range: 0.1 to 300 V in 8 ranges
Dc volts
Accuracy: 3% FS
Range: 0 to 10% and 0 to 30%
Distortion
Accuracy: 1.5% on 0 to 10%
3.5% on 0 to 30
Range: 0 to 20 dB
Sinad
Accuracy: 1.5 dB
Range: 0 to 100%
Am %
Accuracy: 10%
Range: 0 to 100 MΩ
Accuracy: 5% at midscale
not verified below 1 MHz.
2Output
level below -80 dBm is indirectly verified.
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287, and AN/GSM-705. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-
to-one ratio between the standard and TI. Where the four-to-one ratio cannot be met, the
actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories listed in
table 3 are issued as
indicated in
paragraph 4 above and
are used in this
calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range:
10 to 50 W
ARA, Model 757LC (MIS-45845)
AMPLIFIER-POWER SUPPLY
ATTENUATOR
Weinschel, Model 9918-20 dB
(9918-20) (p/o Weinschel, Model
Accuracy: 0.625 dB
9918 set)
AUDIO ANALYZER
Capability: 0.7 to 50% from
Boonton, 1121 (1121)
20 to 9999.9 Hz
Range:
105 to 125 V ac
General Radio, Type
Accuracy: 1%
W10MT3AS3 or Ridge, Model
9020F (9020F)