TB 9-6625-2066-24
5. Accessories Required. The accessories required for the calibration are common usage
accessories, issued as indicated in paragraph 4 above and are not listed in this calibration
procedure.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Fluke, Model 5720A (5720A)
CALIBRATOR
Range: 0 to 306 V at 400 Hz to 300 kHz
(p/o MIS-35947); w amplifier,
Accuracy: +.5%
Fluke 5725A/AR (5725A/AR);
Wideband voltage:
w/ac divider, Fluke, Model 7405A-
Voltage: 1 mV to 3 V
4207 (7405A-4207)
Amplitude flatness: +% output
10 Hz to 1 MHz .5%
1 MHz to 2 MHz 1.25%
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the applicable sections before beginning
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each set within the performance
check where applicable.
CAUTION
Do not apply 115 V ac to TI if batteries are removed.
a. Remove protective cover as required for adjustments.
b. Assure 115 - 230 switch (rear panel) is set to 115.