TB 9-6625-2069-24
Test instrument
parameters
Performance specifications
Period
Range: 20 ns to 999 ms
Accuracy:
(3% of program value + 0.3 ns), <100 ns
(2% of program value), >100 ns
Range: 10 ns to 999 ms1
Width
Accuracy: (1% of program value + 1 ns)
Delay
Range: 0 ns to 999 ms (from 50% point of leading
edge of trigger pulse to 50% point of
leading edge of signal output)
Accuracy: (1% of program value + 1 ns),
D
A SEP B, 50
Range: 20 ns to 999 ms2
ouble pulse
Accuracy: (1% of program value + 1 ns)
Transition times
Range: 6 ns to 9.99 ms LEE and TRE
Accuracy: (3% of program value + 1 ns), A SEP B,
O
50
Range: -9.89 to +9.99 V (HIL, 50 )
utput level
-9.99 TO +9.89 V (LOL, 50 )
-19.7 to +19.9 V (HIL, 1 k )
-19.9 to +19.7 V (LOL, 1 k )
Accuracy: (1% of program value + 1% of amplitude
+50 mV) at 50
(1% of program value + 1% of amplitude
P
+ 100 mV) at 1 k
Preshoot, overshoot, ringing (5% of amplitude
ulse aberration
+ 10 mV)
1Not
calibrated above 900 ms.
2Not
calibrated above 80 ns.
Table 2. Minimum Specifications of Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
Range: 19.1 ns to 1019 s
Fluke, Model PM6681/656
Accuracy: .075%
(PM6681/656)
Range: -20.4 to 20.4 V dc
Fluke, Model 8840A/AF05
Accuracy: 0.05%
(AN/GSM-64D)
Range: 1 ns to 10.2 ms
Agilent OS-303/G (OS-303/G)
Accuracy: ( 3%)
Common name
Description (part number)
CABLE #1
Hewlett-Packard, Model 08160-61610
EXTENDER BOARD
Hewlett-Packard, Model 08160-66573
TERMINATION
Narda, Model 374BNM (374 BNM)
3