Delay/Double Pulse Accuracy
Width Accuracy

Calibration Procedure For Pulse Generator, Model 8160A With Options 001 And 020 Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
. . . .
TB 9-6625-2069-24
NOTE
Cables must be of equal length.
(2) Press pushbuttons and enter data as listed in (a) through (i) below (both
channels for option 020).
PERIOD to 100 ns.
(a)
DELAY to 0 ns.
(b)
HIL to 1 V.
(c)
LOL to 0 V.
(d)
WIDTH to 10 ns.
(e)
LEE to 3 ns.
(f)
TRE to 3 ns.
(g)
BURST to 0.
(h)
INPUT MODE to NORM.
(i)
(3) Measure delay, between trigger output and pulse at 50 percent amplitude as
indicated on oscilloscope (fig. 2). Delay will be  1 ns.
Figure 2. Delay width accuracy.
(4) Repeat technique of (3) above, using DELAY data listed in table 6. Oscilloscope
indication will be within limits specified.
Table 6. Delay - Low Range
Test instrument
Oscilloscope
DELAY
indications
data
(ns)
(ns)
Min
Max
5.4
4.3
6.5
9.6
8.5
10.7
40
38.6
41.4
50
48.5
51.5
99.9 1
97.9
101.9
100
98.0
102.0
PERIOD to 1 s.
1Set
(5) Press DBL PLS pushbutton and enter data of 20 ns.