Transition Time
Table 10. Output Level Accuracy

Calibration Procedure For Pulse Generator, Model 8160A With Options 001 And 020 Page Navigation
  5    6    7    8    9  10  11    12    13    14  
img
. . . . .
TB 9-6625-2069-24
12. Output Level Accuracy
a. Performance Check
(1) Connect equipment as shown in figure 3.
(2) Press pushbuttons and enter data as listed in (a) through (l) below: (Both
channels for option 020).
PERIOD to 999 ms.
(a)
DBL PLS to 20 ns.
(b)
DELAY to 0 ns.
(c)
WIDTH to 10 ns.
(d)
LEE to 3 ns.
(e)
TRE to 3 ns.
(f)
BURST to 0.
(g)
HIL to .1 V.
(h)
LOL to 0.0 V.
(i)
OUTPUT MODE to ENABLE, A SEP B.
(j)
CHANNEL A to COMPL and 50  .
(k)
CHANNEL B to COMPL and 50  (option 020).
(l)
Figure 3. Output - equipment setup.
(3) Multimeter will indicate between 48 and 152 mV.
(4) Repeat technique of (2) and (3) above, using data listed in table 10. Multimeter
will indicate within limits specified.
11