Calibration Procedure For Pulse Generator, Model 8160A With Options 001 And 020 Page Navigation
  9    10    11    12    13  14
img
. . . . . . .
TB 9-6625-2069-24
b. Adjustments
(1) Adjust R95 (fig. 1) for a 5.00 V indication on multimeter (R).
(2) Adjust R50 (fig. 1) for a 20.00 V indication on multimeter (R).
(3) Adjust R71 (fig. 1) for a -20.00 V indication on multimeter (R).
(4) Adjust R223 (fig. 1) for a 20.00 V indication on multimeter (R).
(5) Adjust R243 (fig. 1) f or a -20.00 V indication on multimeter (R).
16. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
15/(16 Blank)