Figure 2. Remote Trigger Level - Equipment Setup.
Figure 5. Remote Trigger Level -2 V.

Calibration Procedure For Digital Electronic Counter, An/Usm-459 (Ts-3662/U) And Models 5328A, 5328Af096 And 5328A041-H69 Page Navigation
  8    9    10    11    12  13  14    15    16  
img
. . . . . . . .
TB 9-6625-2071-24
(7) Adjust oscilloscope Vertical 1
control so that the top of TI marker output
waveform on channel 1 (fig. 3) just barely intersects both positive and negative slopes of
sine wave on channel 2 (fig. 3). If this intersection does not occur at 0 V 35 mV on sine
wave, as shown in fig. 3, perform b below.
Figure 3. Remote Trigger Level Zero.
(8) Move oscilloscope Vertical 1 connection from TI CHANNEL A MARKER
OUTPUT to TI CHANNEL B MARKER OUTPUT.
(9) Repeat (7) above.
(10) Execute TI command string: PF4G6S1S3A379+200*B37+200*R.
(11) Adjust oscilloscope Vertical 1
control so that the top of TI marker output
waveform on channel 1 (fig. 4) just barely intersects both positive and negative slopes of
sine wave on channel 2 (fig. 4). If this intersection does not occur at +2.00 V 35 mV on
sine wave, as shown in fig. 4, perform b below.
Figure 4. Remote Trigger Level +2 V.
(12) Move oscilloscope Vertical 1 connection from TI CHANNEL B MARKER
OUTPUT to TI CHANNEL A MARKER OUTPUT.
(13) Repeat (11) above.
(14) Execute TI command string: PF4G6S1S3A379-200*B37-200*R.