Figure 6. Trigger Level Adjustment - Equipment Setup.

Calibration Procedure For Digital Electronic Counter, An/Usm-459 (Ts-3662/U) And Models 5328A, 5328Af096 And 5328A041-H69 Page Navigation
  11    12    13    14    15  16
img
. . . .
TB 9-6625-2071-24
(18) Adjust resistance standard and function generator offset until multimeter
indicates -2.000 V dc 0.002 V dc.
(19) Adjust oscilloscope controls as necessary to view waveform. Adjust A11R26 (fig. 1)
for a 50 percent duty cycle on oscilloscope display.
(20) Move oscilloscope Vertical 1 connection from TI CHANNEL A MARKER
OUTPUT to TI CHANNEL B MARKER OUTPUT.
(21) Adjust oscilloscope controls as necessary to view waveform. Adjust A11R17 (fig. 1)
for a 50 percent duty cycle on oscilloscope display.
(22) Set function generator Offset to 0 V dc.
14. Final Procedure
a. Deenergize and disconnect all equipment and reinstall TI protective cover.
b. Annotate and affix DA label/form in accordance with TB 750-25.
17/(18 blank)