TB 9-6625-2071-35
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Channel A and B sensitivity:
25 mVrms:
Dc to 40 MHz (dc coupled)
20 Hz to 40 MHz (ac coupled)
50 mVrms:
40 to 100 MHz
AN/USM-459 and 5328AF096
15 mVrms:
Dc to 35 MHz (dc coupled)
20 Hz to 35 MHz (ac coupled)
50 mVrms:
35 to 100 MHz
Channel C sensitivity:
15 mVrms:
5 to 512 MHz
Option 030
15 mVrms:
30 to 500 MHz
AN/USM-459 and 5328AF096
Range: 10, 100, 1000 V dc and auto ranging
Accuracy:
10, 100 V ranges: 0.03% of reading 0.004% of range
1000 V range: 0.087% of reading 0.004% of range
Remote trigger levels:
Range: -2 to +2 V
Accuracy: 35 mV
Option 010, AN/USM-459, and 5328AF096
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI. Where the four-to-one ratio cannot be met, the actual accuracy of the
equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range: 105 to 125 V ac
Ridge, Model 9020A (9020A)
CALIBRATOR
Range: -9.5 to 950 V dc
Fluke, Model 5720A (5700A/EP) (p/o
Accuracy: 0.0085%
MIS-35947); w amplifier, Fluke
5725A/AR (5725A/AR)
Range: 5 part in 1010
Tracor, Model 527E (MIS-10318)
METER
Range: 10 Hz to 1 MHz
(SG-1288/G)
Amplitude: 1 mV to 6 Vp-p
Range: -2 to +2 V dc
Fluke, Model 8840A/AF05
Accuracy: 0.01%
(AN/GSM-64D)
Vertical: 500 mV to 6 Vp-p
(OS-303/G)
Horizontal: 100 Hz to 20 kHz
Accuracy: 3%