Figure 4. Time base and delay generator - right view.

Calibration Procedure For Oscilloscope, Os-189A(P), And Models 180A, 180C And 180D; Dual Channel Vertical Amplifier, Pl-1186/Usm-281 And Pl-1186A/Usm-281A, And, Model 1801A; And Time Base And Delay Generator, Pl-1187/Usm281 And Pl-1187A/Usm-281A, And, Model 1821A Page Navigation
  10    11    12    13    14  15  16    17    18  
img
. . . .
TB 9-6625-2072-24
g. Position TI controls as listed in (1) through (10) below:
(1) MAIN and DELAYED VERNIER controls fully cw (CAL).
(2) DELAYED INT/AUTO/EXT/EXT + 10 switch to INT.
10/EXT/INT LINE switch to INT.
(3) MAIN EXT
(4) MAIN and DELAYED ACS/ACF/AC/DC switches to AC.
(5) TIME/DIV sweep display switch to MAIN.
(6) MAIN TIME/DIV switch to 0.1 s.
(7) SWEEP MODE switch to AUTO.
(8) DELAYED TIME-DIV switch to OFF.
(9) MAIN and DELAYED SLOPE switches to +.
(10) DELAY (DIV) control to 1.00.
21. Sweep Time
a. Performance Check
(1) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to vertical
amplifier A INPUT, using 50  feedthrough termination.
(2) Set oscilloscope calibrator for a 0.1 s MARKER output.
(3) Adjust vertical amplifier VOLTS/DIV controls for approximately 3 divisions of
vertical deflection on oscilloscope.
(4) Adjust MAIN TRIGGER LEVEL control for stable display.
(5) Adjust oscilloscope HORIZONTAL POSITION control to align second marker
with second vertical graticule line. Rotate knob located below EDIT FIELD button on
oscilloscope calibrator to align 10th marker with 10th vertical graticule line. If oscilloscope
calibrator err readout is not within 3 percent, perform b (1) below.
(6) Repeat technique of (2) through (5) above for MAIN TIME/DIV switch settings
listed in table 7. If oscilloscope does not display 1 marker per division between 2d and 10th
vertical graticule lines within 3 percent indication on oscilloscope calibrator err readout,
perform corresponding adjustments listed in table 7.
(7) Set MAIN MIXED DELAYED switch to DELAYED.
(8) Set DELAYED TIME/DIV switch to .1 SEC and MAIN TIME/DIV switch to
.2 SEC.
(9) Repeat technique of (2) through (6) above for DELAYED TIME/DIV switch. If
10th marker is not aligned with 10th vertical graticule line within 3 percent on oscilloscope
calibrator err readout, perform b (2) below.