Table 14. Accuracy
RF Millivoltmeter Accuracy - Continued

Calibration Procedure For Rf Millivoltmeter Me-526/Usm (Model 92Bd With Serial Number 2890 And Below) And Models 92B, 92Bd With Serial Number 2891 And Avobe, And 92C Page Navigation
  16    17    18    19    20  21  22    23  
img
. . . .
TB 9-6625-2074-24
Figure 6. Frequency response - equipment setup.
NOTE
Verify that the proper CAL FACTORS are loaded for the
measuring receiver power sensor module.
(20) Adjust signal generator amplitude controls for a 100.0 mV indication on
measuring receiver. Record TI indication under TI instrument indications No. 1 column in
table 15.
Table 15. Frequency Response
Test instrument indications1
Signal generator
frequency control
No. 1
No. 2
Average
Min
Max
settings (MHz)
(mV)
(mV)
(mV)
(mV)
(mV)
10.00
97
103
30.00
97
103
60.00
97
103
80.00
97
103
100.00
97
103
140.00
97
103
200.00
95
105
400.00
95
105
600.00
95
105
800.00
91
109
1000.00
91
109
1200.00
91
109
'If TI indicates beyond 100 mV range, go to next higher range.
(21) Repeat technique of (19) through (20) above for remaining signal generator
frequency settings listed in table 15.
21