Table 11. Frequency Response - Continued
SECTION V. CALIBRATION PROCESS FOR BOONTON, MODEL 92C

Calibration Procedure For Rf Millivoltmeter Me-526/Usm (Model 92Bd With Serial Number 2890 And Below) And Models 92B, 92Bd With Serial Number 2891 And Avobe, And 92C Page Navigation
  12    13    14    15    16  17  18    19    20    21    22  
img
. . . . . . . .
TB 9-6625-2074-24
(6) Adjust calibrator controls for an output of 30 mV ac and adjust R328 (fig. 5) for a
TI indication between 29.4 and 30.6 mV (ideal 30 mV) (R).
(7) Adjust calibrator controls for an output of 100 mV ac and adjust R341 (fig. 5) for
a TI indication between 98 and 102 mV (ideal 100 mV) (R).
(8) Adjust calibrator controls for an output of 0.3 V ac and adjust R347 (fig. 5) for a
TI indication between 294 and 306 mV (ideal 300 mV) (R).
(9) Adjust calibrator controls for an output of 1.0 V ac and adjust R353 (fig. 5) for a
TI indication between 980 and 1020 mV (ideal 1000 mV) (R).
(10) Adjust calibrator controls for an output of 3.0 V ac and adjust R1405 (fig. 5) for a
TI indication between 2940 and 3060 mV (ideal 3000 mV) (R).
14. Power Supply
NOTE
Do not perform power supply check if all other parameters are
within tolerance.