TB 9-6625-2074-24
Figure 4. Frequency response - equipment setup.
NOTE
Verify that the proper CAL FACTORS are loaded for the measuring
receiver power sensor module.
(27) Zero and calibrate the measuring receiver power sensor module.
(28) Set measuring receiver to measure RF power in mV.
(29) Adjust signal generator frequency controls to frequency setting listed in table 11.
(30) Adjust signal generator amplitude controls for a 100.0 mV indication on
measuring receiver. Record TI indication under TI instrument indications No. 1 column in
table 11.
(31) Repeat technique of (29) through (30) above for remaining signal generator
frequency settings listed in table 11.
Table 11. Frequency Response
Signal generator
Test instrument indications1
frequency control
settings
No. 1
No. 2
Average
Min
Max
(MHz)
(mV)
(mV)
(mV)
(mV)
(mV)
10.00
98
102
30.00
98
102
60.00
98
102
80.00
98
102
100.00
98
102
140.00
98
102
See footnote at end of table.