• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 6. A and B Sweep Accuracy
Power Supply

Calibration Procedure For Oscilloscope, Types 453, 453A, 454, 454Mod163D, And 468 Page Navigation
  11    12    13    14    15  16  17    18    19    20    21  
img
. . . . .
TB 9-6625-2083-24
(3) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for
an Err display of 0.0%.
(4) Adjust C740A (fig. 5) to align marker on 10th vertical graticule line.
(5) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for
an Err display of 0.0%.
(6) Adjust C882 and C892 (fig. 4) for 1 marker 2 divisions (R).
Figure 5. Test instrument - right side view.
15. Risetime
a. Performance Check
(1) Position switches as listed in (a) through (d) below:
CH 1 and CH 2 VOLTS/DIV to 20 mV.
(a)
A AND B TIME/DIV AND DELAY TIME to .1 s.
(b)
MAG to X10.
(c)
HORIZ DISPLAY to A.
(d)
(2) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to CH 1
INPUT, using 50  feedthrough termination.
(3) Set oscilloscope calibrator for a CHAN 1, EDGE mode output of 1  s and
amplitude of 100 mV, and rotate oscilloscope calibrator knob located below EDIT FIELD
pushbutton for 5 divisions of vertical deflection on TI.
(4) Measure risetime, using standard risetime technique. Risetime will be 7 ns or less.
(5) Set TI MODE switch to CH 2.
(6) Set oscilloscope calibrator for a CHAN 2, EDGE mode output of 1  s and
amplitude of 100 mV, and rotate oscilloscope calibrator knob located below EDIT FIELD
pushbutton for 5 divisions of vertical deflection on TI.
(7) Measure risetime, using standard risetime technique. Risetime will be 7 ns or less.
18


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.