TB 9-6625-2083-24
Figure 4. Test instrument - top view.
12. Gain and Magnifier Register
a. Performance Check
(1) Position controls as listed in (a) through (c) below:
(a) A AND B TIME/DIV AND DELAY TIME switches to 1 ms.
(b) HORIZ DISPLAY switch to A.
(c) DELAY-TIME MULTIPLIER dial fully ccw.
(2) Adjust horizontal POSITION control to align second marker on second vertical
graticule line. Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton
for 1 marker per division on TI. If oscilloscope calibrator Err display readout is not within
2%, perform b (1) and (2) below.
(3) Set oscilloscope calibrator for a CHAN 1, MARKER output of .1 ms and set TI
MAG switch to X10.
(4) Repeat technique of (2) above. If TI oscilloscope calibrator Err display readout
is not within 3%, perform b (3) and (4) below.
(5) Set MAG switch to OFF.
(6) Set A AND B TIME/DIV AND DELAY TIME switches to .2 ms and oscilloscope
calibrator MARKER output to 1 ms.
(7) Set MAG switch to X10.
(8) Adjust horizontal POSITION control to align leading edge of middle time
marker to center vertical graticule line. Set MAG switch to OFF. If middle time marker
14