TB 9-6625-2084-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of Oscilloscope, OS-262(P)/U (Tektronix, Type 7623A) and Tektronix, Types
7623, 7613/R7613, R7623A, 7633/R7633, and 7633 Opt 5. The manufacturers' manuals
were used as the prime data sources in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Types 7633/R7633 have reduced span capabilities. Types with
Option 5 operate from 50 to 400 Hz line frequency. OS-262(P)/U is military designation for
Type 7623A.
b. Time and Technique. The time required for this calibration is approximately
2 hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a. Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
115 V 1%, 50 to 60 Hz
Input power requirements
Vertical and horizontal linearity
Range: Dc to 1 MHz
0.5 minor division
Accuracy:
Calibrated sweep rate
Range: 10 ns to 1 ms
0.5 minor division
Accuracy:
Calibrator output
Range: 40 mV, 0.4 and 4.0 V dc
1%
Accuracy:
Risetime
3.5 ns or less
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287, or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.