TB 9-6625-2088-24
(9) Measure symmetry of square wave. If T1 is not within 0.5 percent of T2 (fig. 1),
perform b (5) below.
b. Adjustments
(1) Adjust R12 (fig. 2) for minimum shift.
(2) Adjust R13 (fig. 2) for greater than 5 ms.
(3) Adjust R16 (fig. 2) for best symmetry 0.5 percent. (R)
(4) Repeat a (6) and (7) above due to interaction.
(5) Adjust R35 (fig. 2) for best symmetry 0.5 percent. (R)
a. Performance Check
(1) Adjust ATTENUATION VERNIER control fully ccw and set FREQ/PERIOD
MULTI (Hz/s) switch to 1 M.
(2) Connect oscilloscope to FUNCTION OUT, using X10 attenuator.
(4) Measure risetime and falltime. Risetime and falltime will be less than 20 ns.
b. Adjustments. No adjustments can be made.