TB 9-6625-2089-24
Table 1. Calibration Description - Continued
Test instrument
parameters
Performance specifications
Range:
0.80 μs
Width:
0.05 to 0.10 μs
Risetime:
0.05 to 0.20 μs
Falltime:
0.02 μs of recorded inputs
Accuracy:
RF input power
Range:
+18 dBW to +33 dBW
1.0 dBW
Accuracy:
RF input frequency
Range:
1070 to 1110 MHz
1087 to 1093 MHz, 0.2 MHz, all
Accuracy:
others 0.6 MHz
Range: 1030 MHz fixed, 1030 SWP 5 MHz; 1030 SWP 20 MHz
Accuracy: 1030 MHz fixed and 1030 SWP
5 MHz 0.1 MHz; 1030 SWP 20 MHz
0.3 MHz
RF output power
Range:
-10 dBm
1 dBm
Accuracy:
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287, and AN/GSM-705. Alternate items may be used by the
calibrating activity when the equipment listed in table 2 is not available. The items
selected must be verified to perform satisfactorily prior to use and must bear evidence of
current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio -between the
standard and TI. Where the four-to-one ratio cannot be met, the actual accuracy of the
equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories issued as indicated in paragraph 4 above and are not listed in this
calibration procedure. When necessary, these items may be substituted by equivalent
items, unless specifically prohibited.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range: 105 to 125 V
Ridge, Model 9020A
(9020A)
DIRECTIONAL
Range: 7.0-12.4 GHz
Narda, Model 3095 (3095)
COUPLER
Fluke, Model PM6681/656
0.0005%
(PM6681/656)
Accuracy:
Range: -12.25 to +28.25 V dc
Agilent, Model 3458A (3458A)
Accuracy:
0.025%
Range: Dc to 50 MHz
Agilent, OS-303/G (OS-303/G)
Accuracy: 3%