TB 9-6625-2089-24
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
PEAK POWER
Range: 18 to 33 dBm
Gigatronics, Model 8502A w/sensor
METER
16934 (8902A))
POWER METER
Range: -10 to -20 dBm
Agilent, Model 437B (13440045)
Accuracy: 0.25 dBm
w/power sensor Agilent, Model
8481 or 8482A (13440043)
Pulse width: 0.5 μs
LeCroy, Model 9210MOD200
(9210MOD200) w/plug-ins,
Amplitude: 4.5 to 5.0 V
LeCroy, Models 9211 (9211) and
9215 (9215) (MIS 45839)
SEMICONDUCTOR
Range: 1010 to 1050 MHz
Agilent, Model
DEVICE
423AOPT03 (423AOPT03)
(CRYSTAL DETECTOR)
Range: 1000 to 1100 MHz
Narda, Model 18500B
Power pulse output: 0.5 to 5.0 kW
SYNTHESIZED SIGNAL
Range: 1010 to 1050 MHz
Anritsu, Model 68369NV
GENERATOR
(68369NV)
VARIABLE
Range: 0 to 30 dB
Weinschel, Model AF1171A-
ATTENUATOR NO. 1
69-34 (AF117A-69-34)
VARIABLE
Range: 5 to 90 dB
Narda, Model 2936
ATTENUATOR NO. 2
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name and item identification number as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Additional maintenance information is contained In TM 11-6625-2611-12 and TM 11-6625-
2611-40 for this TI.
d. When indications specified in paragraphs 8 through 21 are not within tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 21. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.